3D Mapping Software Development for 128ch. ECG system
Young Oh Han, Sun Gook Yoo
1Department of Electronic and Information Com. Engineering, Namseoul Univeristy, Korea.
2Department of Biomedical Engineering, Yonsei Univeristy, Korea.
Abstract

The ability to map and ablate the sometimes fleeting automatic atrial tachycardia is greatly enhanced by computerized mapping systems. In this study, we have developed 128 channel computerized data analysis system using microcomputer for basic research of electrophysiology and electrical propagation. The bipolar electrogram information is acquired from 128 cardiac sites simultaneously at a sampling rate of 1 ksampls/sec with continuous and total data storage of up to 30 seconds. When the reference electrogram is selected and reference point is picked up, delay time from the reference point is displayed on three dimensional diagram of the heart. System design permits easy expansion to almost 256 simul- taneous sites. This system is expected to enable us to study pathophysiology of cardiac arrhythmia and to improve the result of diagnosis and surgical treatment for cardiac arrhythmia.

Key Words: 128Ch. ECG System, 3D Mapping, Labview, Opengl
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 0 Crossref
  •    
  • 1,686 View
  • 20 Download
Related articles in Healthc Inform Res


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
FOR CONTRIBUTORS
Editorial Office
1618 Kyungheegung Achim Bldg 3, 34, Sajik-ro 8-gil, Jongno-gu, Seoul 03174, Korea
Tel: +82-2-733-7637, +82-2-734-7637    E-mail: hir@kosmi.org                

Copyright © 2024 by Korean Society of Medical Informatics.

Developed in M2community

Close layer
prev next